Femoral Neck Torsion Angle Measurement by Computed Tomography

Femoral Neck Torsion Angle Measurement by Computed Tomography To establish the accuracy and reproducibility of measuring femoral neck torsion angles (FNTA) on CT images, this angle was measured on 12 dry femora directly on the gross specimens, on plain radiographs, and on CT images. The selection of the CT level, both at the femoral neck and the condyles, that may affect the measurement of FNTA was analyzed. The CT measurements of FNTA were found to have a reproducibility of 2.5 and 0° mean offset error when compared with readings on plain radiographs. A new CT method was developed to measure this angle particularly in patients with coxa valga in whom this measurement is difficult if not impossible to accomplish by current methods. For this purpose composite CT images were made of the femoral head and the femoral shaft, immediately below the lesser trochanter. On these images the line connecting the centers of the femoral head and shaft were inferred to represent the cervical plane. In comparison with the standard CT method, this new method had the same reproducibility, although it slightly overestimated this angle by 2°. This degree of error is negligible when compared with the normal variations of this angle, which ranges from −20 to + 38°. Furthermore, it compares favorably with the limited accuracy of derotational osteotomy surgery. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Journal of Computer Assisted Tomography Wolters Kluwer Health

Femoral Neck Torsion Angle Measurement by Computed Tomography

Loading next page...
 
/lp/wolters-kluwer-health/femoral-neck-torsion-angle-measurement-by-computed-tomography-zd9X5dX9v8

References

References for this paper are not available at this time. We will be adding them shortly, thank you for your patience.

ISSN
0363-8715
eISSN
1532-3145

Abstract

To establish the accuracy and reproducibility of measuring femoral neck torsion angles (FNTA) on CT images, this angle was measured on 12 dry femora directly on the gross specimens, on plain radiographs, and on CT images. The selection of the CT level, both at the femoral neck and the condyles, that may affect the measurement of FNTA was analyzed. The CT measurements of FNTA were found to have a reproducibility of 2.5 and 0° mean offset error when compared with readings on plain radiographs. A new CT method was developed to measure this angle particularly in patients with coxa valga in whom this measurement is difficult if not impossible to accomplish by current methods. For this purpose composite CT images were made of the femoral head and the femoral shaft, immediately below the lesser trochanter. On these images the line connecting the centers of the femoral head and shaft were inferred to represent the cervical plane. In comparison with the standard CT method, this new method had the same reproducibility, although it slightly overestimated this angle by 2°. This degree of error is negligible when compared with the normal variations of this angle, which ranges from −20 to + 38°. Furthermore, it compares favorably with the limited accuracy of derotational osteotomy surgery.

Journal

Journal of Computer Assisted TomographyWolters Kluwer Health

Published: Sep 1, 1987

There are no references for this article.