Correction of single‐beam sample absorption error in a hemispherical 45°/0° spectrophotometer measurement cavity

Correction of single‐beam sample absorption error in a hemispherical 45°/0° spectrophotometer... A mathematical model for quantifying single‐beam sample absorption error in a hemispherical 45°/0° measurement cavity is derived and tested by measuring diffuse reflectance standards with a prototype visible‐range spectrophotometer. A methodology is described for using this three‐parameter model to correct for sample absorption error in any single‐beam spectrophotometer. © 2013 Wiley Periodicals, Inc. Col Res Appl, 39, 436–441, 2014 http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Color Research & Application Wiley

Correction of single‐beam sample absorption error in a hemispherical 45°/0° spectrophotometer measurement cavity

Color Research & Application , Volume 39 (5) – Oct 1, 2014

Loading next page...
 
/lp/wiley/correction-of-single-beam-sample-absorption-error-in-a-hemispherical-ndxd9gXCPY

References (8)

Publisher
Wiley
Copyright
"© 2013 Wiley Periodicals, Inc."
ISSN
0361-2317
eISSN
1520-6378
DOI
10.1002/col.21824
Publisher site
See Article on Publisher Site

Abstract

A mathematical model for quantifying single‐beam sample absorption error in a hemispherical 45°/0° measurement cavity is derived and tested by measuring diffuse reflectance standards with a prototype visible‐range spectrophotometer. A methodology is described for using this three‐parameter model to correct for sample absorption error in any single‐beam spectrophotometer. © 2013 Wiley Periodicals, Inc. Col Res Appl, 39, 436–441, 2014