Evaluation of an iterative model–based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography

Evaluation of an iterative model–based reconstruction algorithm for low-tube-voltage (80 kVp)... Abstract. The objective of this study was to investigate the improvement in diagnostic quality of an iterative model–based reconstruction (IMBR) algorithm for low-tube-voltage (80-kVp) and low-tube-current in abdominal computed tomography angiography (CTA). A total of 11 patients were imaged on a 256-slice multidetector computed tomography for visualization of the aorta. For all patients, three different reconstructions from the low-tube-voltage data are generated: filtered backprojection (FBP), IMBR, and a mixture of both IMBR + FBP . To determine the diagnostic value of IMBR-based reconstructions, the image quality was assessed. With IMBR-based reconstructions, image noise could be significantly reduced, which was confirmed by a highly improved contrast-to-noise ratio. In the image quality assessment, radiologists were able to reliably detect more third-order and higher aortic branches in the IMBR reconstructions compared to FBP reconstructions. The effective dose level was, on average, 3.0 mSv for 80-kVp acquisitions. Low-tube-voltage CTAs significantly improve vascular contrast as presented by others; however, this effect in combination with IMBR enabled yet another substantial improvement of diagnostic quality. For IMBR, a significant improvement of image quality and a decreased radiation dose at low-tube-voltage can be reported. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Journal of Medical Imaging SPIE

Evaluation of an iterative model–based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography

Loading next page...
 
/lp/spie/evaluation-of-an-iterative-model-based-reconstruction-algorithm-for-WJnr1FPFBc
Publisher
SPIE
Copyright
© The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Subject
Physics of Medical Imaging; Paper
ISSN
2329-4302
eISSN
2329-4310
DOI
10.1117/1.JMI.1.3.033501
pmid
26158054
Publisher site
See Article on Publisher Site

Abstract

Abstract. The objective of this study was to investigate the improvement in diagnostic quality of an iterative model–based reconstruction (IMBR) algorithm for low-tube-voltage (80-kVp) and low-tube-current in abdominal computed tomography angiography (CTA). A total of 11 patients were imaged on a 256-slice multidetector computed tomography for visualization of the aorta. For all patients, three different reconstructions from the low-tube-voltage data are generated: filtered backprojection (FBP), IMBR, and a mixture of both IMBR + FBP . To determine the diagnostic value of IMBR-based reconstructions, the image quality was assessed. With IMBR-based reconstructions, image noise could be significantly reduced, which was confirmed by a highly improved contrast-to-noise ratio. In the image quality assessment, radiologists were able to reliably detect more third-order and higher aortic branches in the IMBR reconstructions compared to FBP reconstructions. The effective dose level was, on average, 3.0 mSv for 80-kVp acquisitions. Low-tube-voltage CTAs significantly improve vascular contrast as presented by others; however, this effect in combination with IMBR enabled yet another substantial improvement of diagnostic quality. For IMBR, a significant improvement of image quality and a decreased radiation dose at low-tube-voltage can be reported.

Journal

Journal of Medical ImagingSPIE

Published: Oct 1, 2014