Measurement of Chlamydia pneumoniae-Specific Immunoglobulin A (IgA) Antibodies by the Microimmunofluorescence (MIF) Method: Comparison of Seven Fluorescein-Labeled Anti-Human IgA Conjugates in an In-House MIF Test Using One Commercial MIF and One Enzyme Immunoassay Kit

Measurement of Chlamydia pneumoniae-Specific Immunoglobulin A (IgA) Antibodies by the... For the serological diagnosis of acute Chlamydia pneumoniae infection, the microimmunofluorescence (MIF) test is the most commonly used method and also the "gold standard" for the measurement of immunoglobulin G (IgG) and IgM antibodies. The role of IgA antibodies in diagnosis has not been established. Commercially available fluorescein-labeled anti-human IgA conjugates have not been systematically compared to each other, and this situation may cause considerable variations in IgA results. Therefore, we tested 261 serum samples from 122 patients with pneumonia for IgA antibodies by using six -chain-specific anti-IgA conjugates in our in-house MIF test, one commercial MIF test, and one enzyme immunoassay (EIA). Interfering IgG antibodies were removed with Gullsorb reagent before the measurement of IgA antibodies. Altogether, 14 significant IgA antibody increases in serum samples between the acute phase and the convalescent phase were detected by at least one of the conjugates in the MIF test, while no increases were found in the IgA EIA. Only one patient showed a significant IgA antibody increase with all of the fluorescein-labeled conjugates. Five significant titer changes were detected by at least two conjugates, and in nine instances, the titer increase was detected by one conjugate only. The titer agreement indicated by kappa coefficients was very good or good for all of the fluorescein-labeled conjugates and the EIA with low antibody titers but decreased with increasing titers. http://www.deepdyve.com/assets/images/DeepDyve-Logo-lg.png Clinical and Vaccine Immunology American Society For Microbiology

Measurement of Chlamydia pneumoniae-Specific Immunoglobulin A (IgA) Antibodies by the Microimmunofluorescence (MIF) Method: Comparison of Seven Fluorescein-Labeled Anti-Human IgA Conjugates in an In-House MIF Test Using One Commercial MIF and One Enzyme Immunoassay Kit

Measurement of Chlamydia pneumoniae-Specific Immunoglobulin A (IgA) Antibodies by the Microimmunofluorescence (MIF) Method: Comparison of Seven Fluorescein-Labeled Anti-Human IgA Conjugates in an In-House MIF Test Using One Commercial MIF and One Enzyme Immunoassay Kit

Clinical and Vaccine Immunology , Volume 10 (1): 8 – Jan 1, 2003

Abstract

For the serological diagnosis of acute Chlamydia pneumoniae infection, the microimmunofluorescence (MIF) test is the most commonly used method and also the "gold standard" for the measurement of immunoglobulin G (IgG) and IgM antibodies. The role of IgA antibodies in diagnosis has not been established. Commercially available fluorescein-labeled anti-human IgA conjugates have not been systematically compared to each other, and this situation may cause considerable variations in IgA results. Therefore, we tested 261 serum samples from 122 patients with pneumonia for IgA antibodies by using six -chain-specific anti-IgA conjugates in our in-house MIF test, one commercial MIF test, and one enzyme immunoassay (EIA). Interfering IgG antibodies were removed with Gullsorb reagent before the measurement of IgA antibodies. Altogether, 14 significant IgA antibody increases in serum samples between the acute phase and the convalescent phase were detected by at least one of the conjugates in the MIF test, while no increases were found in the IgA EIA. Only one patient showed a significant IgA antibody increase with all of the fluorescein-labeled conjugates. Five significant titer changes were detected by at least two conjugates, and in nine instances, the titer increase was detected by one conjugate only. The titer agreement indicated by kappa coefficients was very good or good for all of the fluorescein-labeled conjugates and the EIA with low antibody titers but decreased with increasing titers.

Loading next page...
 
/lp/american-society-for-microbiology/measurement-of-chlamydia-pneumoniae-specific-immunoglobulin-a-iga-oE9SpHTRA2

References (33)

Publisher
American Society For Microbiology
Copyright
Copyright © 2003 by the American Society For Microbiology.
ISSN
1556-6811
eISSN
1556-6811
DOI
10.1128/CDLI.10.1.8-12.2003
Publisher site
See Article on Publisher Site

Abstract

For the serological diagnosis of acute Chlamydia pneumoniae infection, the microimmunofluorescence (MIF) test is the most commonly used method and also the "gold standard" for the measurement of immunoglobulin G (IgG) and IgM antibodies. The role of IgA antibodies in diagnosis has not been established. Commercially available fluorescein-labeled anti-human IgA conjugates have not been systematically compared to each other, and this situation may cause considerable variations in IgA results. Therefore, we tested 261 serum samples from 122 patients with pneumonia for IgA antibodies by using six -chain-specific anti-IgA conjugates in our in-house MIF test, one commercial MIF test, and one enzyme immunoassay (EIA). Interfering IgG antibodies were removed with Gullsorb reagent before the measurement of IgA antibodies. Altogether, 14 significant IgA antibody increases in serum samples between the acute phase and the convalescent phase were detected by at least one of the conjugates in the MIF test, while no increases were found in the IgA EIA. Only one patient showed a significant IgA antibody increase with all of the fluorescein-labeled conjugates. Five significant titer changes were detected by at least two conjugates, and in nine instances, the titer increase was detected by one conjugate only. The titer agreement indicated by kappa coefficients was very good or good for all of the fluorescein-labeled conjugates and the EIA with low antibody titers but decreased with increasing titers.

Journal

Clinical and Vaccine ImmunologyAmerican Society For Microbiology

Published: Jan 1, 2003